FIGURE

Fig. 2

ID
ZDB-FIG-200306-65
Publication
Truong et al., 2020 - High-contrast, synchronous volumetric imaging with selective volume illumination microscopy
Other Figures
All Figure Page
Back to All Figure Page
Fig. 2

Higher contrast achieved by SVIM.

a SVIM images of the cranial vasculature improve in contrast as the depth (axial extent) of the illumination volume is decreased. Images are averaged-intensity z-projections of the same 40-µm thick sub-volume, centered at ~170 µm into the head of a 5 dpf zebrafish larva. The SVIM image quality progressively approaches the performance of SPIM as the axial extent of the illumination is reduced to 300 µm or 100 µm, far exceeding the image contrast obtained with wide-field LFM. Inset shows the approximate location of the imaged volume, in context of the zebrafish head. b Quantitative comparison of image contrast, defined as the normalized standard deviation of the pixel values (Methods section), comparing LFM, SPIM, and SVIM of different SVI extents from a. SVIM of smaller extents yielded increasingly better contrast, approaching the performance of SPIM. The contrast of SPIM showed the intrinsic contrast variation of the 3D sample, coupled with the expected contrast decay for increasing imaging depth. The local increase in contrast seen for the SVIM and LFM cases around z = 0 µm came from grid-like artifacts from the light-field reconstruction centered around the native focal plane, a known feature of LFM in general3,4. Scale bars, 100 µm.

Expression Data

Expression Detail
Antibody Labeling
Phenotype Data

Phenotype Detail
Acknowledgments
This image is the copyrighted work of the attributed author or publisher, and ZFIN has permission only to display this image to its users. Additional permissions should be obtained from the applicable author or publisher of the image. Full text @ Commun Biol